For over 40 years, fritted glass tubes have been a staple in thermal desorption sampling. Camsco now offers an innovative, cleaner alternative: Glass ATD Tubes with Internal Restriction and Glass Wool Plugs.
By replacing the fused glass frit with a precisely formed glass restriction, this design creates a stable internal shelf to support your sorbent bed while offering significant operational advantages for high-throughput and demanding analytical applications. This versatile restriction design can also be used without glass wool for direct desorption and solid-phase microextraction (SPME).
Key Advantages
- Consistency & Reusability: Without the porous matrix of a frit—which can trap SVOCs and heavy compounds over time—restriction-based tubes are significantly easier to clean, recondition, and reuse.
- Consistent Flow Profile: Glass frits suffer from manufacturing variances in pore size, permeability, and frit-welding stress, creating flow and pressure drop inconsistencies. A precision glass restriction paired with a glass wool plug minimizes pressure drop variations across tube batches.
- Extended Tube Lifecycle: By simply emptying the tube and replacing the glass wool plug, a restricted tube can be completely refreshed and restored to like-new performance. Even if irreversible contamination with SVOCs occurs, emptying the tube, thermally conditioning the bare glass, and repacking with fresh sorbent and glass wool restores it fully.
- Custom Sorbent Bed Configuration: The glass wool shelf provides a reliable foundation for single- or multi-bed sorbent packing, ensuring stable bed geometry during sample collection and thermal desorption.
- Reduced Active Sites: Combined with Camsco’s proprietary inertness treatments, eliminating the fused glass frit reduces active sites where sensitive target analytes could degrade or adsorb.
Fritted vs. Restriction Comparison
| Feature | Fritted Glass Tubes | Restriction Glass Tubes |
|---|---|---|
| Sorbent Support | Fused Porous Glass Frit | Internal Glass Restriction + Glass Wool Plug |
| Ease of Cleaning | Moderate; frit pores trap VOCs and SVOCs | High; fully renewable by replacing glass wool or thermally conditioning bare tube |
| Pressure Drop Uniformity | Subject to frit porosity and frit-weld variations | Highly Uniform across tube lots |
| Reusability | Standard lifespan | Extended; simple repacking and thermal reconditioning lifecycle |
Specifications & Ordering
- Manufacturer: Camsco (Texas, USA)
- Material: High-purity Borosilicate Glass
- Restriction Locations:
- 15 mm (Standard sorbent support)
- 30 mm (Typically used for direct desorption)
- 8 mm (Engineered specifically for Gerstel TD 3.5+)
- Configurations: Available empty or pre-packed with custom sorbent beds
- Compatibility: Standard industry dimensions compatible with all major automated thermal desorption (ATD) systems